In a previous article, we discussed the basics of the IEPE accelerometers. In this article, we’ll look at the diagram of a typical power unit that can be used to supply power to an IEPE sensor. We’ll also look at the advantages and limitations of the IEPE type and charge output sensors.

Powering IEPE Sensors—Current Regulating Diodes

Figure 1 shows the basic diagram of a power unit that can be used with IEPE sensors.

An example block diagram of a power unit with IEPE sensors.

Figure 1. An example block diagram of a power unit with IEPE sensors. Image used courtesy of Dytran

To begin with, a well-regulated DC power source along with a current regulating circuit is required to supply power to the sensor. The DC power source provides typically between 18 to 30 V. As shown in the Figure, current regulation can be achieved through a current regulating diode (CRD). CRDs are diodes that provide a constant current regardless of voltage fluctuations and load resistance changes. A CRD can keep the current constant over a wide range of voltages from less than 1 V to 100 V. The schematic symbol of a CRD is shown below in Figure 2.

A schematic of a CRD

Figure 2. A schematic of a CRD.

While a majority of the IEPE sensors can be powered by a 2 to 4 mA current source, currents as large as 20 mA are required in some cases. The maximum current rating of CRDs is typically 4 mA. For larger currents, either several CRDs are used in parallel, or a high capacity constant current circuit is used instead.

Note that a power supply that doesn’t have current limiting functionality should not be used to power an IEPE sensor or the sensor will be immediately destroyed.

The IEPE power units normally have a voltmeter that monitors the sensor bias voltage and informs the user of the system status (a short, open, or normal operating condition). Keep in mind that, with an IEPE sensor, both the output signal and the supply voltage are carried through a two-wire connection.

As shown in Figure 1, many IEPE power units use a coupling capacitor to separate the signal from the DC bias voltage. There are also power units that use DC coupling to avoid reducing the discharge time constant of the system.

Typical IEPE Sensor Systems

As shown in Figure 3, there are two options for powering an IEPE sensor. Some readout instruments have built-in constant current power supplies specifically designed for IEPE sensors. If the readout device doesn’t include an IEPE power supply, a separate power unit should be employed, as shown in the Figure.

Options for powering an IEPE sensor.

Figure 3. Options for powering an IEPE sensor. Image used courtesy of PCB Piezotronics

An example readout device that supports IEPE sensors is shown in Figure 4.

An example readout device and diagram for IEPE sensors.

Figure 4. An example readout device and diagram for IEPE sensors. Image used courtesy of NI

Figure 4 shows the NI 9234, which is a four-channel signal acquisition module with software selectable IEPE signal conditioning. As the input circuit diagram shows, the device has a 2 mA constant current power supply and an AC coupling option.

Typical Charge Output (Non-IEPE) Systems

Figure 5 shows a typical measurement system based on a charge output sensor, i.e., a piezoelectric sensor that doesn’t have an integral pre-amplifier.

n example charge output sensor measurement system.

Figure 5. An example charge output sensor measurement system. Image used courtesy of PCB Piezotronics

In this case, an external charge amplifier is required to convert the charge output of the sensor to a voltage signal. In addition, a well-shielded low-noise coaxial cable should be used to connect the sensor to the charge amplifier. However, a standard coaxial cable can be used between the charge amplifier and the readout instrument.

IEPE Sensors vs. Charge Output Sensors

Most applications today use the IEPE type; however, the charge output sensor has its own advantages. Below, we’ll assess the pros and cons of each method.

Sensor Operating Temperature Ranges

The built-in electronics of an IEPE sensor place an upper limit on the operating temperature range of the system. With standard electronics, the maximum temperature is typically about 125˚C. Using special electronic components can push the upper limit to about 165˚C.

On the other hand, a charge output sensor doesn’t have internal electronics, and its high-temperature limit is set by the Curie temperature of the piezoelectric material. The Curie temperature specifies the temperature above which a material loses its piezoelectric properties. Charge output sensors can operate at temperatures as high as 700˚C.

Furthermore, the IEPE type accelerometers can be popular in the general-purpose testing market, while charge output piezoelectric accelerometers dominate applications where temperatures above about 150˚C are involved.

In addition to the Curie temperature limitation, the insulating resistance of the materials employed in a charge output sensor can also limit the maximum operating temperature of the system. In a previous article, we discussed that the sensor insulation resistance can significantly decrease at high temperatures. We also explained that a small insulation resistance can cause a drift in charge amplifiers.

Along with the drift issue, other unexpected and unknown characteristics might be observed when connecting a general-purpose charge amplifier to a sensor with a very small insulation resistance. Figure 6 shows the frequency response of a general-purpose charge amplifier connected to a low resistance sensor.

A graph of the frequency response of a charge amplifier with a low resistance sensor.

Figure 6. A graph of the frequency response of a charge amplifier with a low resistance sensor. Image used courtesy of Endevco

At very low insulation resistance values, the magnitude response of the system exhibits more than 10 dB of peaking at the lower end of the desired frequency range. When working at high temperatures, we need to choose a charge amplifier that can handle very low insulation resistances and maintain a flat frequency response.

Cable Requirements

Charge output sensors require specially treated cables to minimize triboelectric noise. The triboelectric noise refers to the unwanted electric charge produced by rubbing two dissimilar materials together. Flexing or vibration of the cable can cause triboelectric noise. The high impedance output of a non-IEPE sensor can be corrupted by this type of noise and therefore, special low-noise cables are required between the charge output sensor and the charge amplifier. These cables are generally more expensive than standard cables and increase the cost of the non-IEPE system.

Moreover, the high-impedance output of a charge output sensor has limited line driving capability, limiting the cable length between the sensor and charge amplifier to below about 10 meters. On the other hand, standard coaxial cables can be used to transmit the low-impedance output of an IEPE sensor over relatively longer distances without significantly increasing the noise level.

Flexibility in Adjusting Sensor Parameters

With IEPE sensors, the sensor parameters, such as sensitivity and time constant, are fixed at the time of manufacture. We can use external circuits to amplify or attenuate the sensor output but we cannot change the sensitivity or time constant of the sensor. On the other hand, the signal produced by a charge output sensor is processed by an external charge amplifier which can provide us with more flexibility in adjusting parameters such as the system time constant. Special types of charge amplifiers can provide us with very long time constants, extending the low-frequency response of the system.

Additional Features

The IEPE sensors can be operated using relatively cost-effective signal conditioners. Some readout instruments have built-in IEPE power units and can directly connect to an IEPE sensor. If the readout instrument doesn’t have a built-in power unit, we can use an external power unit, as illustrated in Figure 3. On the other hand, with charge output sensors, we need an external charge amplifier. Although laboratory-type charge amplifiers have many helpful features, they are much more expensive than a simple IEPE power unit.

With IEPE sensors, the high impedance signals are placed inside a sealed rugged construction. This makes IEPE sensors less sensitive to harsh conditions such as dirt and humidity.


To see a complete list of my articles, please visit this page.